Article: Inappropriate implantable cardioverter defibrillator shock from a transcutaneous muscle stimulation device therapy
Title | Inappropriate implantable cardioverter defibrillator shock from a transcutaneous muscle stimulation device therapy |
---|---|
Authors | |
Keywords | ICD shock Transcutaneous electrical neuromuscle stimulation |
Issue Date | 2005 |
Publisher | Springer New York LLC. The Journal's web site is located at http://springerlink.metapress.com/openurl.asp?genre=journal&issn=1383-875X |
Citation | Journal Of Interventional Cardiac Electrophysiology, 2005, v. 13 n. 1, p. 73-75 How to Cite? |
Abstract | Inappropriate shock from implantable cardioverter defibrillator (ICD) may result from external electromagnetic interference (EMI), especially for unipolar ventricle sensing. Previous case reports and small in-vitro safety study suggested that endocardial bipolar lead system may be immune from EMI resulting from transcutaneous electrical neuromuscle stimulation (TENS) therapy. This report presents an unusual case of inappropriate discharge in a patient with ICD of endocardial bipolar lead system, receiving TENS from a commercially available device. © 2005 Springer Science + Business Media, Inc. |
Persistent Identifier | http://hdl.handle.net/10722/77090 |
ISSN | 2023 Impact Factor: 2.1 2023 SCImago Journal Rankings: 0.767 |
ISI Accession Number ID | |
References |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Siu, CW | en_HK |
dc.contributor.author | Tse, HF | en_HK |
dc.contributor.author | Lau, CP | en_HK |
dc.date.accessioned | 2010-09-06T07:28:09Z | - |
dc.date.available | 2010-09-06T07:28:09Z | - |
dc.date.issued | 2005 | en_HK |
dc.identifier.citation | Journal Of Interventional Cardiac Electrophysiology, 2005, v. 13 n. 1, p. 73-75 | en_HK |
dc.identifier.issn | 1383-875X | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/77090 | - |
dc.description.abstract | Inappropriate shock from implantable cardioverter defibrillator (ICD) may result from external electromagnetic interference (EMI), especially for unipolar ventricle sensing. Previous case reports and small in-vitro safety study suggested that endocardial bipolar lead system may be immune from EMI resulting from transcutaneous electrical neuromuscle stimulation (TENS) therapy. This report presents an unusual case of inappropriate discharge in a patient with ICD of endocardial bipolar lead system, receiving TENS from a commercially available device. © 2005 Springer Science + Business Media, Inc. | en_HK |
dc.language | eng | en_HK |
dc.publisher | Springer New York LLC. The Journal's web site is located at http://springerlink.metapress.com/openurl.asp?genre=journal&issn=1383-875X | en_HK |
dc.relation.ispartof | Journal of Interventional Cardiac Electrophysiology | en_HK |
dc.subject | ICD shock | - |
dc.subject | Transcutaneous electrical neuromuscle stimulation | - |
dc.subject.mesh | Aged | en_HK |
dc.subject.mesh | Defibrillators, Implantable - adverse effects | en_HK |
dc.subject.mesh | Device Removal | en_HK |
dc.subject.mesh | Electrocardiography | en_HK |
dc.subject.mesh | Electrodes, Implanted | en_HK |
dc.subject.mesh | Electromagnetic Fields - adverse effects | en_HK |
dc.subject.mesh | Equipment Failure | en_HK |
dc.subject.mesh | Follow-Up Studies | en_HK |
dc.subject.mesh | Humans | en_HK |
dc.subject.mesh | Male | en_HK |
dc.subject.mesh | Risk Assessment | en_HK |
dc.subject.mesh | Shock - etiology - physiopathology | en_HK |
dc.subject.mesh | Tachycardia, Ventricular - diagnosis - surgery | en_HK |
dc.subject.mesh | Transcutaneous Electric Nerve Stimulation - adverse effects - methods | en_HK |
dc.title | Inappropriate implantable cardioverter defibrillator shock from a transcutaneous muscle stimulation device therapy | en_HK |
dc.type | Article | en_HK |
dc.identifier.openurl | http://library.hku.hk:4550/resserv?sid=HKU:IR&issn=1383-875X&volume=13&issue=1&spage=73&epage=5&date=2005&atitle=Inappropriate+implantable+cardioverter+defibrillator+shock+from+a+transcutaneous+muscle+stimulation+device+therapy | en_HK |
dc.identifier.email | Siu, CW:cwdsiu@hkucc.hku.hk | en_HK |
dc.identifier.email | Tse, HF:hftse@hkucc.hku.hk | en_HK |
dc.identifier.authority | Siu, CW=rp00534 | en_HK |
dc.identifier.authority | Tse, HF=rp00428 | en_HK |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1007/s10840-005-0357-3 | en_HK |
dc.identifier.pmid | 15976983 | - |
dc.identifier.scopus | eid_2-s2.0-20844443823 | en_HK |
dc.identifier.hkuros | 113806 | en_HK |
dc.relation.references | http://www.scopus.com/mlt/select.url?eid=2-s2.0-20844443823&selection=ref&src=s&origin=recordpage | en_HK |
dc.identifier.volume | 13 | en_HK |
dc.identifier.issue | 1 | en_HK |
dc.identifier.spage | 73 | en_HK |
dc.identifier.epage | 75 | en_HK |
dc.identifier.isi | WOS:000230033100012 | - |
dc.publisher.place | United States | en_HK |
dc.identifier.scopusauthorid | Siu, CW=7006550690 | en_HK |
dc.identifier.scopusauthorid | Tse, HF=7006070805 | en_HK |
dc.identifier.scopusauthorid | Lau, CP=7401968501 | en_HK |
dc.identifier.citeulike | 242103 | - |
dc.identifier.issnl | 1383-875X | - |