File Download
There are no files associated with this item.
Links for fulltext
(May Require Subscription)
- Publisher Website: 10.1093/bja/72.4.388
- Scopus: eid_2-s2.0-0028221837
- PMID: 8155436
- WOS: WOS:A1994NG94700004
- Find via
Supplementary
- Citations:
- Appears in Collections:
Article: Use of two oximeters to investigate a method of movement artefact rejection using photoplethysmographic signals
Title | Use of two oximeters to investigate a method of movement artefact rejection using photoplethysmographic signals |
---|---|
Authors | |
Keywords | Gastrointestinal surgery Hypoxaemia Measurement techniques Monitoring Motion artefact Oximeters |
Issue Date | 1994 |
Publisher | Oxford University Press. The Journal's web site is located at http://bja.oxfordjournals.org/ |
Citation | British Journal Of Anaesthesia, 1994, v. 72 n. 4, p. 388-392 How to Cite? |
Abstract | Oxygen haemoglobin saturations and plethysmograph signal amplitudes were recorded from two oximeters placed on the fingers and toes of 10 patients undergoing oesophagectomy, to assess a method of removing motion artefact from saturation recordings. By examining changes in the plethysmograph amplitude that preceded changes in saturation, episodes of desaturation caused by movement artefacts were removed from the data. The reliability of the method was then determined by scrutinizing two concurrent oximetric profiles from each patient. A total of 1600 h of data were evaluated. Desaturations occurring contemporaneously in both oxygen saturation profiles were presumed genuine, whereas a desaturation occurring in only one of the profiles was classified as artefactual. Our method had a sensitivity of 96%, a positive predictive power of 98% and a specificity of 60%. We modified the method to increase specificity and re-evaluated our data. We found that a useful increase in specificity was associated with a considerable decline in sensitivity. |
Persistent Identifier | http://hdl.handle.net/10722/147168 |
ISSN | 2023 Impact Factor: 9.1 2023 SCImago Journal Rankings: 2.397 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Visram, AR | en_HK |
dc.contributor.author | Jones, RDM | en_HK |
dc.contributor.author | Irwin, MG | en_HK |
dc.contributor.author | BaconShone, J | en_HK |
dc.date.accessioned | 2012-05-29T06:00:32Z | - |
dc.date.available | 2012-05-29T06:00:32Z | - |
dc.date.issued | 1994 | en_HK |
dc.identifier.citation | British Journal Of Anaesthesia, 1994, v. 72 n. 4, p. 388-392 | en_HK |
dc.identifier.issn | 0007-0912 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/147168 | - |
dc.description.abstract | Oxygen haemoglobin saturations and plethysmograph signal amplitudes were recorded from two oximeters placed on the fingers and toes of 10 patients undergoing oesophagectomy, to assess a method of removing motion artefact from saturation recordings. By examining changes in the plethysmograph amplitude that preceded changes in saturation, episodes of desaturation caused by movement artefacts were removed from the data. The reliability of the method was then determined by scrutinizing two concurrent oximetric profiles from each patient. A total of 1600 h of data were evaluated. Desaturations occurring contemporaneously in both oxygen saturation profiles were presumed genuine, whereas a desaturation occurring in only one of the profiles was classified as artefactual. Our method had a sensitivity of 96%, a positive predictive power of 98% and a specificity of 60%. We modified the method to increase specificity and re-evaluated our data. We found that a useful increase in specificity was associated with a considerable decline in sensitivity. | en_HK |
dc.language | eng | en_US |
dc.publisher | Oxford University Press. The Journal's web site is located at http://bja.oxfordjournals.org/ | en_HK |
dc.relation.ispartof | British Journal of Anaesthesia | en_HK |
dc.subject | Gastrointestinal surgery | en_HK |
dc.subject | Hypoxaemia | en_HK |
dc.subject | Measurement techniques | en_HK |
dc.subject | Monitoring | en_HK |
dc.subject | Motion artefact | en_HK |
dc.subject | Oximeters | en_HK |
dc.subject.mesh | Aged | en_US |
dc.subject.mesh | Aged, 80 And Over | en_US |
dc.subject.mesh | Esophagectomy | en_US |
dc.subject.mesh | Humans | en_US |
dc.subject.mesh | Male | en_US |
dc.subject.mesh | Middle Aged | en_US |
dc.subject.mesh | Movement | en_US |
dc.subject.mesh | Oximetry - Methods | en_US |
dc.subject.mesh | Oxygen - Blood | en_US |
dc.subject.mesh | Photoplethysmography | en_US |
dc.subject.mesh | Sensitivity And Specificity | en_US |
dc.subject.mesh | Time Factors | en_US |
dc.title | Use of two oximeters to investigate a method of movement artefact rejection using photoplethysmographic signals | en_HK |
dc.type | Article | en_HK |
dc.identifier.email | Irwin, MG: mgirwin@hku.hk | en_HK |
dc.identifier.email | BaconShone, J: johnbs@hku.hk | en_HK |
dc.identifier.authority | Irwin, MG=rp00390 | en_HK |
dc.identifier.authority | BaconShone, J=rp00056 | en_HK |
dc.description.nature | link_to_subscribed_fulltext | en_US |
dc.identifier.doi | 10.1093/bja/72.4.388 | - |
dc.identifier.pmid | 8155436 | - |
dc.identifier.scopus | eid_2-s2.0-0028221837 | en_HK |
dc.identifier.volume | 72 | en_HK |
dc.identifier.issue | 4 | en_HK |
dc.identifier.spage | 388 | en_HK |
dc.identifier.epage | 392 | en_HK |
dc.identifier.isi | WOS:A1994NG94700004 | - |
dc.publisher.place | United Kingdom | en_HK |
dc.identifier.scopusauthorid | Visram, AR=6602931233 | en_HK |
dc.identifier.scopusauthorid | Jones, RDM=7501542949 | en_HK |
dc.identifier.scopusauthorid | Irwin, MG=7202411076 | en_HK |
dc.identifier.scopusauthorid | BaconShone, J=6602137416 | en_HK |
dc.identifier.issnl | 0007-0912 | - |