File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: Prone table stereotactic breast biopsy

TitleProne table stereotactic breast biopsy
Authors
KeywordsBiopsy
Breast neoplasms
Mammography
Stereotaxic techniques, instrumentation
Issue Date2002
PublisherHong Kong Medical Association. The Journal's web site is located at http://www.hkmj.org/resources/supp.html
Citation
Hong Kong Medical Journal, 2002, v. 8 n. 6, p. 447-451 How to Cite?
AbstractThe prone table machine is a mammographic X-ray system specially designed for use in the stereotactic localisation of breast abnormality. In this study, its clinical usefulness was investigated in terms of duration, success rate, complications, and patients' acceptance of the procedure. During a 5-month period, 79 patients attended the Kwong Wah Hospital for stereotactic-guided biopsy on the prone table. Eighty-one lesions were assessed - seven by fine needle aspirations, 67 by large-core needle biopsies, and seven by vacuum-assisted biopsies. Most of the biopsies were done because of clustered microcalcifications (77.8%) and the majority were of mammographically indeterminate nature (58.0%). The mean duration of the procedure was 49 minutes. A high degree of acceptance was experienced by patients. Only one patient had persistent haemorrhage after the biopsy. In conclusion, the prone table machine was considered to be useful and efficient, and had a high degree of acceptance among patients.
Persistent Identifierhttp://hdl.handle.net/10722/154220
ISSN
2021 Impact Factor: 1.256
2020 SCImago Journal Rankings: 0.357
References

 

DC FieldValueLanguage
dc.contributor.authorHui, JYHen_US
dc.contributor.authorChan, LKen_US
dc.contributor.authorChan, RLMen_US
dc.contributor.authorLau, AWLen_US
dc.contributor.authorLo, Jen_US
dc.contributor.authorChan, JCSen_US
dc.contributor.authorLam, HSen_US
dc.date.accessioned2012-08-08T08:23:57Z-
dc.date.available2012-08-08T08:23:57Z-
dc.date.issued2002en_US
dc.identifier.citationHong Kong Medical Journal, 2002, v. 8 n. 6, p. 447-451en_US
dc.identifier.issn1024-2708en_US
dc.identifier.urihttp://hdl.handle.net/10722/154220-
dc.description.abstractThe prone table machine is a mammographic X-ray system specially designed for use in the stereotactic localisation of breast abnormality. In this study, its clinical usefulness was investigated in terms of duration, success rate, complications, and patients' acceptance of the procedure. During a 5-month period, 79 patients attended the Kwong Wah Hospital for stereotactic-guided biopsy on the prone table. Eighty-one lesions were assessed - seven by fine needle aspirations, 67 by large-core needle biopsies, and seven by vacuum-assisted biopsies. Most of the biopsies were done because of clustered microcalcifications (77.8%) and the majority were of mammographically indeterminate nature (58.0%). The mean duration of the procedure was 49 minutes. A high degree of acceptance was experienced by patients. Only one patient had persistent haemorrhage after the biopsy. In conclusion, the prone table machine was considered to be useful and efficient, and had a high degree of acceptance among patients.en_US
dc.languageengen_US
dc.publisherHong Kong Medical Association. The Journal's web site is located at http://www.hkmj.org/resources/supp.htmlen_US
dc.relation.ispartofHong Kong Medical Journalen_US
dc.subjectBiopsy-
dc.subjectBreast neoplasms-
dc.subjectMammography-
dc.subjectStereotaxic techniques, instrumentation-
dc.subject.meshAdulten_US
dc.subject.meshAgeden_US
dc.subject.meshBiopsy, Needle - Instrumentation - Methodsen_US
dc.subject.meshBreast - Pathologyen_US
dc.subject.meshBreast Neoplasms - Pathology - Radiographyen_US
dc.subject.meshCalcinosis - Radiographyen_US
dc.subject.meshFemaleen_US
dc.subject.meshHumansen_US
dc.subject.meshMammography - Instrumentation - Methodsen_US
dc.subject.meshMiddle Ageden_US
dc.subject.meshPatient Satisfactionen_US
dc.subject.meshProne Positionen_US
dc.subject.meshSensitivity And Specificityen_US
dc.subject.meshTime Factorsen_US
dc.titleProne table stereotactic breast biopsyen_US
dc.typeArticleen_US
dc.identifier.emailLo, J:drjohnlo@hku.hken_US
dc.identifier.authorityLo, J=rp00041en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.pmid12459602-
dc.identifier.scopuseid_2-s2.0-0036900488en_US
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0036900488&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume8en_US
dc.identifier.issue6en_US
dc.identifier.spage447en_US
dc.identifier.epage451en_US
dc.publisher.placeHong Kongen_US
dc.identifier.scopusauthoridHui, JYH=35080369800en_US
dc.identifier.scopusauthoridChan, LK=34871581500en_US
dc.identifier.scopusauthoridChan, RLM=36907987900en_US
dc.identifier.scopusauthoridLau, AWL=36947588700en_US
dc.identifier.scopusauthoridLo, J=7201650902en_US
dc.identifier.scopusauthoridChan, JCS=15724681800en_US
dc.identifier.scopusauthoridLam, HS=7202774979en_US
dc.identifier.issnl1024-2708-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats