File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: The performance of three vacuum-mixing cement guns - A comparison of the fatigue properties of Simplex P cement

TitleThe performance of three vacuum-mixing cement guns - A comparison of the fatigue properties of Simplex P cement
Authors
Issue Date2001
PublisherSpringer. The Journal's web site is located at http://www.springer.com/medicine/orthopedics/journal/264
Citation
International Orthopaedics, 2001, v. 25 n. 5, p. 290-293 How to Cite?
AbstractWe tested three commercially available cement-mixing systems (Howmedica, Summit and Stryker). After mixing the cement was allowed to cure in metal moulds under standard conditions. Sixty-nine cement blocks were produced. The fatigue behavior of the blocks was tested until failure. A wide scattering of results was observed in all the three systems. Uneven mixing of the polymer and monomer was observed in the Howmedica system. A significantly stronger cement (P=0.018) was produced with the Stryker system.
Persistent Identifierhttp://hdl.handle.net/10722/170029
ISSN
2023 Impact Factor: 2.0
2023 SCImago Journal Rankings: 0.877
ISI Accession Number ID
References

 

DC FieldValueLanguage
dc.contributor.authorYau, WPen_US
dc.contributor.authorNg, TPen_US
dc.contributor.authorChiu, KYen_US
dc.contributor.authorPoon, KCen_US
dc.contributor.authorHo, WYen_US
dc.contributor.authorLuk, DKen_US
dc.date.accessioned2012-10-30T06:04:50Z-
dc.date.available2012-10-30T06:04:50Z-
dc.date.issued2001en_US
dc.identifier.citationInternational Orthopaedics, 2001, v. 25 n. 5, p. 290-293en_US
dc.identifier.issn0341-2695en_US
dc.identifier.urihttp://hdl.handle.net/10722/170029-
dc.description.abstractWe tested three commercially available cement-mixing systems (Howmedica, Summit and Stryker). After mixing the cement was allowed to cure in metal moulds under standard conditions. Sixty-nine cement blocks were produced. The fatigue behavior of the blocks was tested until failure. A wide scattering of results was observed in all the three systems. Uneven mixing of the polymer and monomer was observed in the Howmedica system. A significantly stronger cement (P=0.018) was produced with the Stryker system.en_US
dc.languageengen_US
dc.publisherSpringer. The Journal's web site is located at http://www.springer.com/medicine/orthopedics/journal/264en_US
dc.relation.ispartofInternational Orthopaedicsen_US
dc.subject.meshBone Cementsen_US
dc.subject.meshChemistry, Physicalen_US
dc.subject.meshHumansen_US
dc.subject.meshMaterials Testing - Methodsen_US
dc.subject.meshOrthopedic Procedures - Instrumentationen_US
dc.subject.meshPhysicochemical Phenomenaen_US
dc.subject.meshSensitivity And Specificityen_US
dc.subject.meshStress, Mechanicalen_US
dc.subject.meshSurface Propertiesen_US
dc.subject.meshTensile Strengthen_US
dc.subject.meshVacuumen_US
dc.titleThe performance of three vacuum-mixing cement guns - A comparison of the fatigue properties of Simplex P cementen_US
dc.typeArticleen_US
dc.identifier.emailYau, WP:peterwpy@hkucc.hku.hken_US
dc.identifier.emailChiu, KY:pkychiu@hkucc.hku.hken_US
dc.identifier.emailLuk, DK:hcm21000@hku.hken_US
dc.identifier.authorityYau, WP=rp00500en_US
dc.identifier.authorityChiu, KY=rp00379en_US
dc.identifier.authorityLuk, DK=rp00333en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1007/s002640100266en_US
dc.identifier.pmid11794261-
dc.identifier.scopuseid_2-s2.0-0034751946en_US
dc.identifier.hkuros65203-
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0034751946&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume25en_US
dc.identifier.issue5en_US
dc.identifier.spage290en_US
dc.identifier.epage293en_US
dc.identifier.isiWOS:000171975500008-
dc.publisher.placeGermanyen_US
dc.identifier.scopusauthoridYau, WP=7005822441en_US
dc.identifier.scopusauthoridNg, TP=24438193400en_US
dc.identifier.scopusauthoridChiu, KY=7202988127en_US
dc.identifier.scopusauthoridPoon, KC=36776891500en_US
dc.identifier.scopusauthoridHo, WY=36855638100en_US
dc.identifier.scopusauthoridLuk, DK=7201921573en_US
dc.identifier.issnl0341-2695-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats