Article: Metachronous multiple non-anastomotic aneurysms in a bifurcated double velour knitted Dacron graft

TitleMetachronous multiple non-anastomotic aneurysms in a bifurcated double velour knitted Dacron graft
Authors
KeywordsDacron Graft
Non-Anastomotic Aneurysms
Issue Date2002
PublisherBlackwell Publishing Asia. The Journal's web site is located at http://www.blackwellpublishing.com/journals/ASH
Citation
Annals Of The College Of Surgeons Of Hong Kong, 2002, v. 6 n. 4, p. 123-125 How to Cite?
AbstractA case of metachronous non-anastomotic aneurysms in a bifurcated double velour knitted Dacron aortobifemoral graft is reported. The cause of non-anastomotic graft aneurysms is most likely due to a basic structural failure in the graft, as a result of defects in fabrication, Dacron fibre deterioration or a combination of both. Local excision of the graft defect with interposition graft replacement is an accepted treatment for local graft failure. However, as fabrication defects and Dacron fibre deterioration are both factors affecting the whole graft, it is not surprising to find multiple aneurysms in the same graft synchronously or metachronously. Thus, a total graft replacement should be the treatment of choice for graft failure.
Persistent Identifierhttp://hdl.handle.net/10722/172810
ISSN
References

 

DC FieldValueLanguage
dc.contributor.authorTing, ACWen_US
dc.contributor.authorHo, Pen_US
dc.contributor.authorCheng, SWKen_US
dc.date.accessioned2012-10-30T06:25:02Z-
dc.date.available2012-10-30T06:25:02Z-
dc.date.issued2002en_US
dc.identifier.citationAnnals Of The College Of Surgeons Of Hong Kong, 2002, v. 6 n. 4, p. 123-125en_US
dc.identifier.issn1028-4001en_US
dc.identifier.urihttp://hdl.handle.net/10722/172810-
dc.description.abstractA case of metachronous non-anastomotic aneurysms in a bifurcated double velour knitted Dacron aortobifemoral graft is reported. The cause of non-anastomotic graft aneurysms is most likely due to a basic structural failure in the graft, as a result of defects in fabrication, Dacron fibre deterioration or a combination of both. Local excision of the graft defect with interposition graft replacement is an accepted treatment for local graft failure. However, as fabrication defects and Dacron fibre deterioration are both factors affecting the whole graft, it is not surprising to find multiple aneurysms in the same graft synchronously or metachronously. Thus, a total graft replacement should be the treatment of choice for graft failure.en_US
dc.languageengen_US
dc.publisherBlackwell Publishing Asia. The Journal's web site is located at http://www.blackwellpublishing.com/journals/ASHen_US
dc.relation.ispartofAnnals of the College of Surgeons of Hong Kongen_US
dc.subjectDacron Graften_US
dc.subjectNon-Anastomotic Aneurysmsen_US
dc.titleMetachronous multiple non-anastomotic aneurysms in a bifurcated double velour knitted Dacron graften_US
dc.typeArticleen_US
dc.identifier.emailCheng, SWK: wkcheng@hkucc.hku.hken_US
dc.identifier.authorityCheng, SWK=rp00374en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1046/j.1442-2034.2002.00147.xen_US
dc.identifier.scopuseid_2-s2.0-0036863035en_US
dc.identifier.hkuros77047-
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0036863035&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume6en_US
dc.identifier.issue4en_US
dc.identifier.spage123en_US
dc.identifier.epage125en_US
dc.publisher.placeAustraliaen_US
dc.identifier.scopusauthoridTing, ACW=7102858552en_US
dc.identifier.scopusauthoridHo, P=24469553100en_US
dc.identifier.scopusauthoridCheng, SWK=7404684779en_US
dc.identifier.issnl1028-4001-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats