Article: A splint for controlled active motion after flexor tendon repair. Design, mechanical testing, and preliminary clinical results
Title | A splint for controlled active motion after flexor tendon repair. Design, mechanical testing, and preliminary clinical results |
---|---|
Authors | |
Issue Date | 1990 |
Publisher | WB Saunders Co. The Journal's web site is located at http://www.elsevier.com/locate/jhsa |
Citation | Journal Of Hand Surgery, 1990, v. 15 n. 4, p. 645-651 How to Cite? |
Abstract | A splint for controlled active motion after flexor tendon repair is described. It incorporates a single core-coated elastic band passing around a palmar pulley and attached proximally to a spring wire. Its mechanical properties were tested against six other systems. The tension in various systems all rose near full extension. However, the palmar pulley, the spring wire, and the elastic band each could lower the tension significantly. When the bending moments at the interphalangeal joints were measured, all systems produced a peak during the latter part of extension. With the palmar pulley, spring wire, and elastic band, the rise was minimal and in fact, the bending moments diminished near full extension. Initial results in 28 flexor tendon repairs using this splint showed less flexion contracture when compared with 78 flexor tendon repairs using a standard rubber band anchored at the wrist. |
Persistent Identifier | http://hdl.handle.net/10722/178176 |
ISSN | 2023 Impact Factor: 2.1 2023 SCImago Journal Rankings: 0.983 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Chow, SP | en_US |
dc.contributor.author | Stephens, MM | en_US |
dc.contributor.author | Ngai, WK | en_US |
dc.contributor.author | Pun, WK | en_US |
dc.contributor.author | Crosby So, CYC | en_US |
dc.contributor.author | Chu, M | en_US |
dc.date.accessioned | 2012-12-19T09:43:15Z | - |
dc.date.available | 2012-12-19T09:43:15Z | - |
dc.date.issued | 1990 | en_US |
dc.identifier.citation | Journal Of Hand Surgery, 1990, v. 15 n. 4, p. 645-651 | en_US |
dc.identifier.issn | 0363-5023 | en_US |
dc.identifier.uri | http://hdl.handle.net/10722/178176 | - |
dc.description.abstract | A splint for controlled active motion after flexor tendon repair is described. It incorporates a single core-coated elastic band passing around a palmar pulley and attached proximally to a spring wire. Its mechanical properties were tested against six other systems. The tension in various systems all rose near full extension. However, the palmar pulley, the spring wire, and the elastic band each could lower the tension significantly. When the bending moments at the interphalangeal joints were measured, all systems produced a peak during the latter part of extension. With the palmar pulley, spring wire, and elastic band, the rise was minimal and in fact, the bending moments diminished near full extension. Initial results in 28 flexor tendon repairs using this splint showed less flexion contracture when compared with 78 flexor tendon repairs using a standard rubber band anchored at the wrist. | en_US |
dc.language | eng | en_US |
dc.publisher | WB Saunders Co. The Journal's web site is located at http://www.elsevier.com/locate/jhsa | en_US |
dc.relation.ispartof | Journal of Hand Surgery | en_US |
dc.subject.mesh | Biomechanics | en_US |
dc.subject.mesh | Contracture - Prevention & Control | en_US |
dc.subject.mesh | Equipment Design | en_US |
dc.subject.mesh | Fingers - Physiopathology - Surgery | en_US |
dc.subject.mesh | Hand - Physiopathology - Surgery | en_US |
dc.subject.mesh | Humans | en_US |
dc.subject.mesh | Movement | en_US |
dc.subject.mesh | Postoperative Period | en_US |
dc.subject.mesh | Splints | en_US |
dc.subject.mesh | Tendons - Physiopathology - Surgery | en_US |
dc.title | A splint for controlled active motion after flexor tendon repair. Design, mechanical testing, and preliminary clinical results | en_US |
dc.type | Article | en_US |
dc.identifier.email | Chow, SP: spchow@hku.hk | en_US |
dc.identifier.authority | Chow, SP=rp00064 | en_US |
dc.description.nature | link_to_subscribed_fulltext | en_US |
dc.identifier.doi | 10.1016/S0363-5023(09)90030-2 | - |
dc.identifier.pmid | 2380531 | - |
dc.identifier.scopus | eid_2-s2.0-0025296667 | en_US |
dc.identifier.volume | 15 | en_US |
dc.identifier.issue | 4 | en_US |
dc.identifier.spage | 645 | en_US |
dc.identifier.epage | 651 | en_US |
dc.identifier.isi | WOS:A1990DP02600020 | - |
dc.publisher.place | United States | en_US |
dc.identifier.scopusauthorid | Chow, SP=7201828376 | en_US |
dc.identifier.scopusauthorid | Stephens, MM=7201575245 | en_US |
dc.identifier.scopusauthorid | Ngai, WK=6602875630 | en_US |
dc.identifier.scopusauthorid | Pun, WK=7003726073 | en_US |
dc.identifier.scopusauthorid | Crosby So, CYC=6504103502 | en_US |
dc.identifier.scopusauthorid | Chu, M=7402456181 | en_US |
dc.identifier.issnl | 0363-5023 | - |