Title | Analysis of electrode condition for whole cell recording |
---|---|
Authors | |
Issue Date | 2013 |
Citation | IEEE International Conference on Nano/Molecular Medicine and Engineering, NANOMED, 2013, p. 43-46 How to Cite? |
Abstract | In this paper, an experimental study on the successful rate of using different types of pipettes for whole cell recording is reported. 5 types of pipettes have been designed and made. The procedure of whole cell recording has been analyzed, and the seal formation of and the rupture of patched membrane have been monitored. According to our research, the optimized value of pipette resistance is 15-30 Mω for whole cell recording. © 2013 IEEE. |
Persistent Identifier | http://hdl.handle.net/10722/213405 |
ISSN | 2020 SCImago Journal Rankings: 0.111 |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Yang, Runhuai | - |
dc.contributor.author | Lai, King W C | - |
dc.contributor.author | Xi, Ning | - |
dc.contributor.author | Yang, Jie | - |
dc.date.accessioned | 2015-07-28T04:07:10Z | - |
dc.date.available | 2015-07-28T04:07:10Z | - |
dc.date.issued | 2013 | - |
dc.identifier.citation | IEEE International Conference on Nano/Molecular Medicine and Engineering, NANOMED, 2013, p. 43-46 | - |
dc.identifier.issn | 2159-6964 | - |
dc.identifier.uri | http://hdl.handle.net/10722/213405 | - |
dc.description.abstract | In this paper, an experimental study on the successful rate of using different types of pipettes for whole cell recording is reported. 5 types of pipettes have been designed and made. The procedure of whole cell recording has been analyzed, and the seal formation of and the rupture of patched membrane have been monitored. According to our research, the optimized value of pipette resistance is 15-30 Mω for whole cell recording. © 2013 IEEE. | - |
dc.language | eng | - |
dc.relation.ispartof | IEEE International Conference on Nano/Molecular Medicine and Engineering, NANOMED | - |
dc.title | Analysis of electrode condition for whole cell recording | - |
dc.type | Conference_Paper | - |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1109/NANOMED.2013.6766313 | - |
dc.identifier.scopus | eid_2-s2.0-84898830732 | - |
dc.identifier.spage | 43 | - |
dc.identifier.epage | 46 | - |
dc.identifier.eissn | 2159-6972 | - |
dc.identifier.issnl | 2159-6964 | - |