Article: Evaluation of a disposable device for patient-controlled epidural analgesia after caesarean section

TitleEvaluation of a disposable device for patient-controlled epidural analgesia after caesarean section
Authors
KeywordsEquipment: epidural, patient controlled analgesia
Pain: postoperative
Analgesics: pethidine
Issue Date1996
Citation
Anaesthesia and Intensive Care, 1996, v. 24, n. 1, p. 51-55 How to Cite?
AbstractWe evaluated a disposable device (Baxter PCA Infusor) for patient-controlled epidural analgesia (PCEA) using pethidine in twenty women after lower segment caesarean section. Efficacy, as measured by visual analogue pain scores, was comparable with historical controls from PCEA studies using electronic devices, Three patients reported inadequate analgesia, attributable in two cases to problems with the epidural catheter PCEA was stopped in one patient because of side-effects. Pethidine consumption ranged from 125 to 1500 mg (median 575 mg) in 48 hours. Plasma concentrations of pethidine varied widely, Disposable devices for PCEA after caesarean section provide an alternative to bolus administration or PCEA using more expensive and cumbersome electronic devices, although we suggest currently available apparatus requires modifications to improve clinical performance.
Persistent Identifierhttp://hdl.handle.net/10722/280456
ISSN
2023 Impact Factor: 1.1
2023 SCImago Journal Rankings: 0.534
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorNgan Kee, W. D.-
dc.contributor.authorLam, K. K.-
dc.contributor.authorTwyford, C. J.-
dc.contributor.authorGin, T.-
dc.date.accessioned2020-02-17T14:34:05Z-
dc.date.available2020-02-17T14:34:05Z-
dc.date.issued1996-
dc.identifier.citationAnaesthesia and Intensive Care, 1996, v. 24, n. 1, p. 51-55-
dc.identifier.issn0310-057X-
dc.identifier.urihttp://hdl.handle.net/10722/280456-
dc.description.abstractWe evaluated a disposable device (Baxter PCA Infusor) for patient-controlled epidural analgesia (PCEA) using pethidine in twenty women after lower segment caesarean section. Efficacy, as measured by visual analogue pain scores, was comparable with historical controls from PCEA studies using electronic devices, Three patients reported inadequate analgesia, attributable in two cases to problems with the epidural catheter PCEA was stopped in one patient because of side-effects. Pethidine consumption ranged from 125 to 1500 mg (median 575 mg) in 48 hours. Plasma concentrations of pethidine varied widely, Disposable devices for PCEA after caesarean section provide an alternative to bolus administration or PCEA using more expensive and cumbersome electronic devices, although we suggest currently available apparatus requires modifications to improve clinical performance.-
dc.languageeng-
dc.relation.ispartofAnaesthesia and Intensive Care-
dc.subjectEquipment: epidural, patient controlled analgesia-
dc.subjectPain: postoperative-
dc.subjectAnalgesics: pethidine-
dc.titleEvaluation of a disposable device for patient-controlled epidural analgesia after caesarean section-
dc.typeArticle-
dc.description.naturelink_to_OA_fulltext-
dc.identifier.doi10.1177/0310057X9602400109-
dc.identifier.pmid8669655-
dc.identifier.scopuseid_2-s2.0-0030029551-
dc.identifier.volume24-
dc.identifier.issue1-
dc.identifier.spage51-
dc.identifier.epage55-
dc.identifier.isiWOS:A1996TW75000010-
dc.identifier.issnl0310-057X-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats