File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: The normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor

TitleThe normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor
Authors
Keywordsdiagnostic techniques and procedures
ultrasonography
hemodynamics
Doppler
pediatrics
resuscitation
Issue Date2010
Citation
Critical Care Medicine, 2010, v. 38, n. 9, p. 1875-1881 How to Cite?
AbstractObjective: The Ultrasonic Cardiac Output Monitor is a noninvasive method of hemodynamic assessment and monitoring in critically ill patients. There are no published reference ranges for normal values in children for this device. This study aimed to establish normal ranges for cardiovascular indices measured using Ultrasonic Cardiac Output Monitor in children aged 0-12 yrs old and to assess interobserver reliability. Design: This was a population-based cross-sectional observational study. Setting: Schools and kindergartens in Hong Kong. Subjects: Chinese children aged up to 12 yrs old. Interventions: Two operators performed Ultrasonic Cardiac Output Monitor scans on each child together with standard oscillometric measurement of blood pressure and heart rate. Software intrinsic to the Ultrasonic Cardiac Output Monitor device produces values for stroke volume, cardiac output, and systemic vascular resistance. For each parameter, normal ranges were defined as lying between the 2.5th and 97.5th percentiles. Interobserver reliability was assessed with Bland-Altman plots, coefficients of variation, and intraclass correlation. Measurements and main results: A total of 1,197 Chinese children (55% boys) were scanned. Normal ranges of values for cardiac output, stroke volume, and systemic vascular resistance indices are presented. Interobserver reliability for Ultrasonic Cardiac Output Monitor was superior to that for standard blood pressure and heart rate measurement. Conclusions: This large study presents normal values for cardiovascular indices in children using the Ultrasonic Cardiac Output Monitor with good interobserver reliability. © 2010 by the Society of Critical Care Medicine and Lippincott Williams & Wilkins.
Persistent Identifierhttp://hdl.handle.net/10722/291985
ISSN
2021 Impact Factor: 9.296
2020 SCImago Journal Rankings: 3.002
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorCattermole, Giles N.-
dc.contributor.authorLeung, P. Y.Mia-
dc.contributor.authorMak, Paulina S.K.-
dc.contributor.authorChan, Stewart S.W.-
dc.contributor.authorGraham, Colin A.-
dc.contributor.authorRainer, Timothy H.-
dc.date.accessioned2020-11-17T14:55:31Z-
dc.date.available2020-11-17T14:55:31Z-
dc.date.issued2010-
dc.identifier.citationCritical Care Medicine, 2010, v. 38, n. 9, p. 1875-1881-
dc.identifier.issn0090-3493-
dc.identifier.urihttp://hdl.handle.net/10722/291985-
dc.description.abstractObjective: The Ultrasonic Cardiac Output Monitor is a noninvasive method of hemodynamic assessment and monitoring in critically ill patients. There are no published reference ranges for normal values in children for this device. This study aimed to establish normal ranges for cardiovascular indices measured using Ultrasonic Cardiac Output Monitor in children aged 0-12 yrs old and to assess interobserver reliability. Design: This was a population-based cross-sectional observational study. Setting: Schools and kindergartens in Hong Kong. Subjects: Chinese children aged up to 12 yrs old. Interventions: Two operators performed Ultrasonic Cardiac Output Monitor scans on each child together with standard oscillometric measurement of blood pressure and heart rate. Software intrinsic to the Ultrasonic Cardiac Output Monitor device produces values for stroke volume, cardiac output, and systemic vascular resistance. For each parameter, normal ranges were defined as lying between the 2.5th and 97.5th percentiles. Interobserver reliability was assessed with Bland-Altman plots, coefficients of variation, and intraclass correlation. Measurements and main results: A total of 1,197 Chinese children (55% boys) were scanned. Normal ranges of values for cardiac output, stroke volume, and systemic vascular resistance indices are presented. Interobserver reliability for Ultrasonic Cardiac Output Monitor was superior to that for standard blood pressure and heart rate measurement. Conclusions: This large study presents normal values for cardiovascular indices in children using the Ultrasonic Cardiac Output Monitor with good interobserver reliability. © 2010 by the Society of Critical Care Medicine and Lippincott Williams & Wilkins.-
dc.languageeng-
dc.relation.ispartofCritical Care Medicine-
dc.subjectdiagnostic techniques and procedures-
dc.subjectultrasonography-
dc.subjecthemodynamics-
dc.subjectDoppler-
dc.subjectpediatrics-
dc.subjectresuscitation-
dc.titleThe normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.1097/CCM.0b013e3181e8adee-
dc.identifier.pmid20562697-
dc.identifier.scopuseid_2-s2.0-77956231366-
dc.identifier.volume38-
dc.identifier.issue9-
dc.identifier.spage1875-
dc.identifier.epage1881-
dc.identifier.eissn1530-0293-
dc.identifier.isiWOS:000281559500015-
dc.identifier.issnl0090-3493-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats