File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: Coronary artery calcium measurement with multi-detector row CT: In vitro assessment of effect of radiation dose

TitleCoronary artery calcium measurement with multi-detector row CT: In vitro assessment of effect of radiation dose
Authors
KeywordsComputed tomography (CT), radiation exposure
Coronary vessels, calcification
Coronary vessels, CT
Issue Date2002
Citation
Radiology, 2002, v. 225, n. 3, p. 901-906 How to Cite?
AbstractThe authors assessed in vitro the effect of radiation dose on coronary artery calcium quantification with multi-detector row computed tomography. A cardiac phantom with calcified cylinders was scanned at various milliampere second settings (20-160 mAs). A clear tendency was found for image noise to decrease as tube current increased (P < .001). No tendency was found for the Agatson score or calcium volume and mass errors to vary with tube current. Calcium measurements were not significantly affected by the choice of tube current. Calcium mass error was strongly correlated with calcium volume error (P < .001). The calcium mass measurement was more accurate and less variable than the calcium volume measurement. © RSNA, 2002.
Persistent Identifierhttp://hdl.handle.net/10722/315926
ISSN
2021 Impact Factor: 29.146
2020 SCImago Journal Rankings: 3.118
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorHong, Cheng-
dc.contributor.authorBae, Kyongtae T.-
dc.contributor.authorPilgram, Thomas K.-
dc.contributor.authorSuh, Jongdae-
dc.contributor.authorBradley, David-
dc.date.accessioned2022-08-24T15:48:37Z-
dc.date.available2022-08-24T15:48:37Z-
dc.date.issued2002-
dc.identifier.citationRadiology, 2002, v. 225, n. 3, p. 901-906-
dc.identifier.issn0033-8419-
dc.identifier.urihttp://hdl.handle.net/10722/315926-
dc.description.abstractThe authors assessed in vitro the effect of radiation dose on coronary artery calcium quantification with multi-detector row computed tomography. A cardiac phantom with calcified cylinders was scanned at various milliampere second settings (20-160 mAs). A clear tendency was found for image noise to decrease as tube current increased (P < .001). No tendency was found for the Agatson score or calcium volume and mass errors to vary with tube current. Calcium measurements were not significantly affected by the choice of tube current. Calcium mass error was strongly correlated with calcium volume error (P < .001). The calcium mass measurement was more accurate and less variable than the calcium volume measurement. © RSNA, 2002.-
dc.languageeng-
dc.relation.ispartofRadiology-
dc.subjectComputed tomography (CT), radiation exposure-
dc.subjectCoronary vessels, calcification-
dc.subjectCoronary vessels, CT-
dc.titleCoronary artery calcium measurement with multi-detector row CT: In vitro assessment of effect of radiation dose-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.1148/radiol.2253011871-
dc.identifier.pmid12461277-
dc.identifier.scopuseid_2-s2.0-0036891972-
dc.identifier.volume225-
dc.identifier.issue3-
dc.identifier.spage901-
dc.identifier.epage906-
dc.identifier.isiWOS:000179420800039-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats