Article: The impact of implant-related characteristics on dental implant blooming: An in vitro study
Title | The impact of implant-related characteristics on dental implant blooming: An in vitro study |
---|---|
Authors | |
Keywords | artifacts blooming cone-beam computed tomography dental implants imaging inaccuracy volumetric alteration |
Issue Date | 2022 |
Citation | Clinical Oral Implants Research, 2022, v. 33, n. 12, p. 1199-1211 How to Cite? |
Abstract | Aim: To assess, in vitro, variables potentially influencing implant blooming using a human-like imaging phantom and 3D-printed mandibles. Material and Methods: Sixty implants were inserted in 3D-printed mandibles in 26 different configurations in order to examine the impact of implant diameter, presence of a cover screw, implant design/material, implant position, and the presence of additional implants on implant blooming using two cone-beam computed tomography (CBCT) devices (Accuitomo [ACC] and NewTom [NWT]). Two observers measured the amount of implant blooming in both buccolingual and mesiodistal directions. Inter-rater agreement and descriptive statistics, grouped by implant characteristic and CBCT device, were calculated. Results: Both CBCT devices increased implant diameter (a mean increase of 9.2% and 11.8% for titanium, 20.3% and 24.4% for zirconium, for ACC and NWT, respectively). An increase in implant diameter did not increase the amount of blooming, whereas placing a cover screw did (from 8.0% to 10.9% for ACC, and from 10.0% to 15.6% for NWT). Moreover, implant design, anatomical region, and the presence of another implant also affected the extent of the blooming. Conclusions: Dental implants show a clear diameter increase on CBCT, with the effect being more pronounced for zirconium than for titanium implants. Similar effects are likely to occur in the clinical setting, potentially masking nonosseointegration, reducing the dimensions of peri-implant defects, and/or causing underestimation of the buccal bone thickness. |
Persistent Identifier | http://hdl.handle.net/10722/323494 |
ISSN | 2023 Impact Factor: 4.8 2023 SCImago Journal Rankings: 1.865 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Tarce, Mihai | - |
dc.contributor.author | de Greef, Alexander | - |
dc.contributor.author | Lahoud, Pierre | - |
dc.contributor.author | de Faria Vasconcelos, Karla | - |
dc.contributor.author | Jacobs, Reinhilde | - |
dc.contributor.author | Quirynen, Marc | - |
dc.date.accessioned | 2023-01-06T14:10:11Z | - |
dc.date.available | 2023-01-06T14:10:11Z | - |
dc.date.issued | 2022 | - |
dc.identifier.citation | Clinical Oral Implants Research, 2022, v. 33, n. 12, p. 1199-1211 | - |
dc.identifier.issn | 0905-7161 | - |
dc.identifier.uri | http://hdl.handle.net/10722/323494 | - |
dc.description.abstract | Aim: To assess, in vitro, variables potentially influencing implant blooming using a human-like imaging phantom and 3D-printed mandibles. Material and Methods: Sixty implants were inserted in 3D-printed mandibles in 26 different configurations in order to examine the impact of implant diameter, presence of a cover screw, implant design/material, implant position, and the presence of additional implants on implant blooming using two cone-beam computed tomography (CBCT) devices (Accuitomo [ACC] and NewTom [NWT]). Two observers measured the amount of implant blooming in both buccolingual and mesiodistal directions. Inter-rater agreement and descriptive statistics, grouped by implant characteristic and CBCT device, were calculated. Results: Both CBCT devices increased implant diameter (a mean increase of 9.2% and 11.8% for titanium, 20.3% and 24.4% for zirconium, for ACC and NWT, respectively). An increase in implant diameter did not increase the amount of blooming, whereas placing a cover screw did (from 8.0% to 10.9% for ACC, and from 10.0% to 15.6% for NWT). Moreover, implant design, anatomical region, and the presence of another implant also affected the extent of the blooming. Conclusions: Dental implants show a clear diameter increase on CBCT, with the effect being more pronounced for zirconium than for titanium implants. Similar effects are likely to occur in the clinical setting, potentially masking nonosseointegration, reducing the dimensions of peri-implant defects, and/or causing underestimation of the buccal bone thickness. | - |
dc.language | eng | - |
dc.relation.ispartof | Clinical Oral Implants Research | - |
dc.subject | artifacts | - |
dc.subject | blooming | - |
dc.subject | cone-beam computed tomography | - |
dc.subject | dental implants | - |
dc.subject | imaging | - |
dc.subject | inaccuracy | - |
dc.subject | volumetric alteration | - |
dc.title | The impact of implant-related characteristics on dental implant blooming: An in vitro study | - |
dc.type | Article | - |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1111/clr.14002 | - |
dc.identifier.pmid | 36189488 | - |
dc.identifier.scopus | eid_2-s2.0-85139621716 | - |
dc.identifier.volume | 33 | - |
dc.identifier.issue | 12 | - |
dc.identifier.spage | 1199 | - |
dc.identifier.epage | 1211 | - |
dc.identifier.eissn | 1600-0501 | - |
dc.identifier.isi | WOS:000866104100001 | - |