Article: Analysis of defects in ProTaper hand-operated instruments after clinical use
Title | Analysis of defects in ProTaper hand-operated instruments after clinical use |
---|---|
Authors | |
Keywords | Clinical defect ProTaper |
Issue Date | 2007 |
Publisher | Elsevier Inc. The Journal's web site is located at http://www.jendodon.com |
Citation | Journal of Endodontics, 2007, v. 33 n. 3, p. 287-290 How to Cite? |
Abstract | The purpose of this study was to analyze the type and location of defects observed in ProTaper for Hand Use (PHU) instruments after routine clinical use. We analyzed a total of 401 PHUs discarded from an endodontic clinic over a 17-month period. Those failed instruments were examined on the lateral and fractographic surface by scanning electron microscope. Of the 86 PHUs that showed discernible defects, 28 were intact but partially unwound, and 58 were fractured (36 because of shear and 22 from fatigue failure). The primary characteristic of shear failure was the presence of a skewed dimple and/or tear ridge, a typical pattern developed because of a combination of various loads. Nearly 74% of the instruments with defects exhibited shear damage. About three-quarters of the instrument fractures occurred in the apical one-third of the canal, mostly in molars. The results of this study indicated that most PHU instruments fail because of either shear or fatigue. © 2007 American Association of Endodontists. |
Persistent Identifier | http://hdl.handle.net/10722/66355 |
ISSN | 2023 Impact Factor: 3.5 2023 SCImago Journal Rankings: 1.356 |
ISI Accession Number ID | |
References |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Shen, Y | en_HK |
dc.contributor.author | Bian, Z | en_HK |
dc.contributor.author | Cheung, GSP | en_HK |
dc.contributor.author | Peng, B | en_HK |
dc.date.accessioned | 2010-09-06T05:45:40Z | - |
dc.date.available | 2010-09-06T05:45:40Z | - |
dc.date.issued | 2007 | en_HK |
dc.identifier.citation | Journal of Endodontics, 2007, v. 33 n. 3, p. 287-290 | en_HK |
dc.identifier.issn | 0099-2399 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/66355 | - |
dc.description.abstract | The purpose of this study was to analyze the type and location of defects observed in ProTaper for Hand Use (PHU) instruments after routine clinical use. We analyzed a total of 401 PHUs discarded from an endodontic clinic over a 17-month period. Those failed instruments were examined on the lateral and fractographic surface by scanning electron microscope. Of the 86 PHUs that showed discernible defects, 28 were intact but partially unwound, and 58 were fractured (36 because of shear and 22 from fatigue failure). The primary characteristic of shear failure was the presence of a skewed dimple and/or tear ridge, a typical pattern developed because of a combination of various loads. Nearly 74% of the instruments with defects exhibited shear damage. About three-quarters of the instrument fractures occurred in the apical one-third of the canal, mostly in molars. The results of this study indicated that most PHU instruments fail because of either shear or fatigue. © 2007 American Association of Endodontists. | en_HK |
dc.language | eng | en_HK |
dc.publisher | Elsevier Inc. The Journal's web site is located at http://www.jendodon.com | en_HK |
dc.relation.ispartof | Journal of Endodontics | en_HK |
dc.rights | The Journal of Endodontics. Copyright © Lippincott Williams & Wilkins. | en_HK |
dc.subject | Clinical | - |
dc.subject | defect | - |
dc.subject | ProTaper | - |
dc.subject.mesh | Dental Alloys | en_HK |
dc.subject.mesh | Dental Instruments | en_HK |
dc.subject.mesh | Dental Stress Analysis | en_HK |
dc.subject.mesh | Equipment Failure Analysis | en_HK |
dc.subject.mesh | Humans | en_HK |
dc.subject.mesh | Microscopy, Electron, Scanning | en_HK |
dc.subject.mesh | Nickel | en_HK |
dc.subject.mesh | Root Canal Preparation - instrumentation | en_HK |
dc.subject.mesh | Shear Strength | en_HK |
dc.subject.mesh | Titanium | en_HK |
dc.title | Analysis of defects in ProTaper hand-operated instruments after clinical use | en_HK |
dc.type | Article | en_HK |
dc.identifier.openurl | http://library.hku.hk:4550/resserv?sid=HKU:IR&issn=0099-2399&volume=33&issue=3&spage=287&epage=290&date=2007&atitle=Analysis+of+defects+in+ProTaper+hand-operated+instruments+after+clinical+use | en_HK |
dc.identifier.email | Cheung, GSp:spcheung@hkucc.hku.hk | en_HK |
dc.identifier.authority | Cheung, GSp=rp00016 | en_HK |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1016/j.joen.2006.09.009 | en_HK |
dc.identifier.pmid | 17320716 | - |
dc.identifier.scopus | eid_2-s2.0-33847081737 | en_HK |
dc.identifier.hkuros | 126613 | en_HK |
dc.relation.references | http://www.scopus.com/mlt/select.url?eid=2-s2.0-33847081737&selection=ref&src=s&origin=recordpage | en_HK |
dc.identifier.volume | 33 | en_HK |
dc.identifier.issue | 3 | en_HK |
dc.identifier.spage | 287 | en_HK |
dc.identifier.epage | 290 | en_HK |
dc.identifier.isi | WOS:000244586500018 | - |
dc.publisher.place | United States | en_HK |
dc.identifier.scopusauthorid | Shen, Y=7404767171 | en_HK |
dc.identifier.scopusauthorid | Bian, Z=7103314705 | en_HK |
dc.identifier.scopusauthorid | Cheung, GSP=7005809531 | en_HK |
dc.identifier.scopusauthorid | Peng, B=7102873002 | en_HK |
dc.identifier.issnl | 0099-2399 | - |