Article: Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation-assisted total knee arthroplasty
Title | Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation-assisted total knee arthroplasty |
---|---|
Authors | |
Keywords | computer assisted orthopedic surgery interobserver and intra-observer errors navigation registration total knee arthroplasty |
Issue Date | 2007 |
Publisher | Churchill Livingstone. The Journal's web site is located at http://www.elsevier.com/locate/arthroplastyjournal |
Citation | Journal of Arthroplasty, 2007, v. 22 n. 8, p. 1150-1161 How to Cite? |
Abstract | This study investigated the errors of obtaining visually selected anatomic landmarks for use in the registration process in a passive optical non-image-based computer-assisted total knee arthroplasty system in 5 fresh frozen cadavers. The projected maximum errors in the femoral mechanical axis (due to registration errors of the center of the distal femur) were 0.7 degrees in the coronal and 1.4 degrees in the sagittal plane. The projected maximum errors in the tibial mechanical axis arising from registration errors of the center of the proximal tibia were 1.3 degrees in the coronal and 2 degrees in the sagittal plane. The projected maximum errors in the transepicondylar axis were 9.1 degrees (registration errors of the medial femoral epicondyle) and 7.2 degrees (registration errors of the lateral femoral epicondyle). It should be noted that the results may be partly related to the use of the particular system in this experiment. |
Persistent Identifier | http://hdl.handle.net/10722/79706 |
ISSN | 2023 Impact Factor: 3.4 2023 SCImago Journal Rankings: 1.849 |
ISI Accession Number ID | |
References |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Yau, WP | en_HK |
dc.contributor.author | Leung, A | en_HK |
dc.contributor.author | Liu, KG | en_HK |
dc.contributor.author | Yan, CH | en_HK |
dc.contributor.author | Wong, LLS | en_HK |
dc.contributor.author | Chiu, KY | en_HK |
dc.date.accessioned | 2010-09-06T07:57:40Z | - |
dc.date.available | 2010-09-06T07:57:40Z | - |
dc.date.issued | 2007 | en_HK |
dc.identifier.citation | Journal of Arthroplasty, 2007, v. 22 n. 8, p. 1150-1161 | en_HK |
dc.identifier.issn | 0883-5403 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/79706 | - |
dc.description.abstract | This study investigated the errors of obtaining visually selected anatomic landmarks for use in the registration process in a passive optical non-image-based computer-assisted total knee arthroplasty system in 5 fresh frozen cadavers. The projected maximum errors in the femoral mechanical axis (due to registration errors of the center of the distal femur) were 0.7 degrees in the coronal and 1.4 degrees in the sagittal plane. The projected maximum errors in the tibial mechanical axis arising from registration errors of the center of the proximal tibia were 1.3 degrees in the coronal and 2 degrees in the sagittal plane. The projected maximum errors in the transepicondylar axis were 9.1 degrees (registration errors of the medial femoral epicondyle) and 7.2 degrees (registration errors of the lateral femoral epicondyle). It should be noted that the results may be partly related to the use of the particular system in this experiment. | en_HK |
dc.language | eng | en_HK |
dc.publisher | Churchill Livingstone. The Journal's web site is located at http://www.elsevier.com/locate/arthroplastyjournal | en_HK |
dc.relation.ispartof | Journal of Arthroplasty | en_HK |
dc.subject | computer assisted orthopedic surgery | - |
dc.subject | interobserver and intra-observer errors | - |
dc.subject | navigation | - |
dc.subject | registration | - |
dc.subject | total knee arthroplasty | - |
dc.subject.mesh | Arthroplasty, Replacement, Knee - methods | - |
dc.subject.mesh | Cadaver | - |
dc.subject.mesh | Surgery, Computer-Assisted - methods | - |
dc.subject.mesh | Tibia - anatomy and histology | - |
dc.subject.mesh | Tomography, X-Ray Computed | - |
dc.title | Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation-assisted total knee arthroplasty | en_HK |
dc.type | Article | en_HK |
dc.identifier.openurl | http://library.hku.hk:4550/resserv?sid=HKU:IR&issn=0883-5403&volume=22&issue=8&spage=1150&epage=1161&date=2007&atitle=Interobserver+and+intra-observer+errors+in+obtaining+visually+selected+anatomical+landmarks+during+registration+process+in+non-image-based+navigation-assisted+total+knee+arthroplasty | en_HK |
dc.identifier.email | Yau, WP: peterwpy@hkucc.hku.hk | en_HK |
dc.identifier.email | Yan, CH: yanchunhoi@yahoo.com.hk | en_HK |
dc.identifier.email | Chiu, KY: pkychiu@hkucc.hku.hk | en_HK |
dc.identifier.authority | Yau, WP=rp00500 | en_HK |
dc.identifier.authority | Yan, CH=rp00303 | en_HK |
dc.identifier.authority | Chiu, KY=rp00379 | en_HK |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1016/j.arth.2006.10.010 | en_HK |
dc.identifier.pmid | 18078884 | - |
dc.identifier.scopus | eid_2-s2.0-36849081392 | en_HK |
dc.identifier.hkuros | 139421 | en_HK |
dc.identifier.hkuros | 205300 | - |
dc.relation.references | http://www.scopus.com/mlt/select.url?eid=2-s2.0-36849081392&selection=ref&src=s&origin=recordpage | en_HK |
dc.identifier.volume | 22 | en_HK |
dc.identifier.issue | 8 | en_HK |
dc.identifier.spage | 1150 | en_HK |
dc.identifier.epage | 1161 | en_HK |
dc.identifier.eissn | 1532-8406 | - |
dc.identifier.isi | WOS:000205636600012 | - |
dc.publisher.place | United States | en_HK |
dc.identifier.scopusauthorid | Chiu, KY=7202988127 | en_HK |
dc.identifier.scopusauthorid | Wong, LLS=21741631900 | en_HK |
dc.identifier.scopusauthorid | Yan, CH=35764094100 | en_HK |
dc.identifier.scopusauthorid | Liu, KG=14016424300 | en_HK |
dc.identifier.scopusauthorid | Leung, A=23035440900 | en_HK |
dc.identifier.scopusauthorid | Yau, WP=7005822441 | en_HK |
dc.identifier.issnl | 0883-5403 | - |