Article: Fractured prosthetic abutments in osseointegrated implants: A technical complication to cope with

TitleFractured prosthetic abutments in osseointegrated implants: A technical complication to cope with
Authors
KeywordsFracture
ITI® Dental Implant System
Oral implants
Prosthetic abutment
Repair
Technical complication
Issue Date2000
PublisherWiley-Blackwell Publishing, Inc.. The Journal's web site is located at http://www.blackwellpublishing.com/journals/CLR
Citation
Clinical Oral Implants Research, 2000, v. 11 n. 2, p. 163-170 How to Cite?
AbstractIn the present report a repair set to retrieve fractured abutments remaining in the implants of the ITI® Dental Implant System is presented. In 1 case in which such fractures occurred, the repair set was successfully applied. After removal of all particles of the fractured prosthetic abutments and recutting the threads, new abutments were used, and the original suprastructure could be reinserted. The causes for such technical failures are discussed considering the relatively sparse information in the literature on that topic. © Munksgaard 2000.
Persistent Identifierhttp://hdl.handle.net/10722/154104
ISSN
2021 Impact Factor: 5.021
2020 SCImago Journal Rankings: 2.407
ISI Accession Number ID
References

 

DC FieldValueLanguage
dc.contributor.authorLuterbacher, Sen_US
dc.contributor.authorFourmousis, Ien_US
dc.contributor.authorLang, NPen_US
dc.contributor.authorBrägger, Uen_US
dc.date.accessioned2012-08-08T08:23:17Z-
dc.date.available2012-08-08T08:23:17Z-
dc.date.issued2000en_US
dc.identifier.citationClinical Oral Implants Research, 2000, v. 11 n. 2, p. 163-170en_US
dc.identifier.issn0905-7161en_US
dc.identifier.urihttp://hdl.handle.net/10722/154104-
dc.description.abstractIn the present report a repair set to retrieve fractured abutments remaining in the implants of the ITI® Dental Implant System is presented. In 1 case in which such fractures occurred, the repair set was successfully applied. After removal of all particles of the fractured prosthetic abutments and recutting the threads, new abutments were used, and the original suprastructure could be reinserted. The causes for such technical failures are discussed considering the relatively sparse information in the literature on that topic. © Munksgaard 2000.en_US
dc.languageengen_US
dc.publisherWiley-Blackwell Publishing, Inc.. The Journal's web site is located at http://www.blackwellpublishing.com/journals/CLRen_US
dc.relation.ispartofClinical Oral Implants Researchen_US
dc.subjectFracture-
dc.subjectITI® Dental Implant System-
dc.subjectOral implants-
dc.subjectProsthetic abutment-
dc.subjectRepair-
dc.subjectTechnical complication-
dc.subject.meshDental Abutmentsen_US
dc.subject.meshDental Implantsen_US
dc.subject.meshDental Prosthesis Repairen_US
dc.subject.meshDental Restoration Failureen_US
dc.subject.meshDenture, Partial, Fixeden_US
dc.subject.meshDevice Removal - Methodsen_US
dc.subject.meshHumansen_US
dc.subject.meshMiddle Ageden_US
dc.titleFractured prosthetic abutments in osseointegrated implants: A technical complication to cope withen_US
dc.typeArticleen_US
dc.identifier.emailLang, NP:nplang@hkucc.hku.hken_US
dc.identifier.authorityLang, NP=rp00031en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1034/j.1600-0501.2000.011002163.x-
dc.identifier.pmid11168207-
dc.identifier.scopuseid_2-s2.0-0034173188en_US
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0034173188&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume11en_US
dc.identifier.issue2en_US
dc.identifier.spage163en_US
dc.identifier.epage170en_US
dc.identifier.isiWOS:000086076700009-
dc.publisher.placeUnited Statesen_US
dc.identifier.scopusauthoridLuterbacher, S=36957661100en_US
dc.identifier.scopusauthoridFourmousis, I=6602718088en_US
dc.identifier.scopusauthoridLang, NP=7201577367en_US
dc.identifier.scopusauthoridBrägger, U=7005538598en_US
dc.identifier.issnl0905-7161-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats