Article: The Crooked Rod Sign: A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure
Title | The Crooked Rod Sign: A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure |
---|---|
Authors | |
Keywords | Complication Deformed threads Distraction failure mechanism Magnetically controlled growing rod Rod slippage |
Issue Date | 2020 |
Publisher | Lippincott, Williams & Wilkins. The Journal's web site is located at http://journals.lww.com/spinejournal/pages/default.aspx |
Citation | Spine, 2020, v. 45 n. 6, p. E346-E351 How to Cite? |
Abstract | STUDY DESIGN: Case report OBJECTIVE.: To report a unique mechanical failure of magnetically controlled growing rods (MCGR) that is related to continuous rod slippage.
SUMMARY OF BACKGROUND DATA: Despite the growing popularity of MCGR in the management of patients with early onset scoliosis, postoperative complications and reoperations are not uncommon. Unique complications or unplanned reoperations are observed in patients with MCGRs that are not seen in traditional growing rods. The complications include rod slippage, mismatch between targeted and achieved distraction length, metallosis and actuator pin fracture. We have identified an unreported failure mechanism whereby deformed threads occur in the internal distraction mechanism of the MCGR. This phenomenon may indicate increased MCGR internal screw friction and increased distraction resistance. This complication ultimately led to distraction failure and revision surgery.
METHODS: A girl with early onset scoliosis (EOS) was treated by dual MCGRs. She developed proximal junctional kyphosis (PJK) and continuous rod slippage. Distractions in the out-patient clinic and under sedation did not achieve successful rod lengthening. A tell-tale 'crooked rod' radiological sign was identified by angulation between the actuator and the extendable portion of the rod.
RESULTS: Rod exchange was performed and she is now 2 years after revision surgery with successful lengthening episodes. Upon review of the extracted MCGR, distraction was not possible even after rod removal and was only resumed after the screw was manually reinserted along its correct threads.
CONCLUSIONS: Deformed threads is a complication that cannot be reverted without rod removal and exchange. Users should be aware of this potential failure mechanism and proceed to early rod exchange rather than attempting any further distractions.
LEVEL OF EVIDENCE: 4.
This is a report of a mechanical failure with magnetically controlled growing rods resulting from continuous distraction attempts despite persistent rod slippage. The internal screw's threads were deformed and the magnetically controlled growing rods could no longer be lengthened. Distraction failure and rod exchange was necessary to proceed to further lengthening. |
Persistent Identifier | http://hdl.handle.net/10722/279699 |
ISSN | 2023 Impact Factor: 2.6 2023 SCImago Journal Rankings: 1.221 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Cheung, JPY | - |
dc.contributor.author | Zhang, T | - |
dc.contributor.author | Bow, HYC | - |
dc.contributor.author | Kwan, KYH | - |
dc.contributor.author | Sze, KY | - |
dc.contributor.author | Cheung, KMC | - |
dc.date.accessioned | 2019-12-09T06:44:36Z | - |
dc.date.available | 2019-12-09T06:44:36Z | - |
dc.date.issued | 2020 | - |
dc.identifier.citation | Spine, 2020, v. 45 n. 6, p. E346-E351 | - |
dc.identifier.issn | 0362-2436 | - |
dc.identifier.uri | http://hdl.handle.net/10722/279699 | - |
dc.description.abstract | STUDY DESIGN: Case report OBJECTIVE.: To report a unique mechanical failure of magnetically controlled growing rods (MCGR) that is related to continuous rod slippage. SUMMARY OF BACKGROUND DATA: Despite the growing popularity of MCGR in the management of patients with early onset scoliosis, postoperative complications and reoperations are not uncommon. Unique complications or unplanned reoperations are observed in patients with MCGRs that are not seen in traditional growing rods. The complications include rod slippage, mismatch between targeted and achieved distraction length, metallosis and actuator pin fracture. We have identified an unreported failure mechanism whereby deformed threads occur in the internal distraction mechanism of the MCGR. This phenomenon may indicate increased MCGR internal screw friction and increased distraction resistance. This complication ultimately led to distraction failure and revision surgery. METHODS: A girl with early onset scoliosis (EOS) was treated by dual MCGRs. She developed proximal junctional kyphosis (PJK) and continuous rod slippage. Distractions in the out-patient clinic and under sedation did not achieve successful rod lengthening. A tell-tale 'crooked rod' radiological sign was identified by angulation between the actuator and the extendable portion of the rod. RESULTS: Rod exchange was performed and she is now 2 years after revision surgery with successful lengthening episodes. Upon review of the extracted MCGR, distraction was not possible even after rod removal and was only resumed after the screw was manually reinserted along its correct threads. CONCLUSIONS: Deformed threads is a complication that cannot be reverted without rod removal and exchange. Users should be aware of this potential failure mechanism and proceed to early rod exchange rather than attempting any further distractions. LEVEL OF EVIDENCE: 4. This is a report of a mechanical failure with magnetically controlled growing rods resulting from continuous distraction attempts despite persistent rod slippage. The internal screw's threads were deformed and the magnetically controlled growing rods could no longer be lengthened. Distraction failure and rod exchange was necessary to proceed to further lengthening. | - |
dc.language | eng | - |
dc.publisher | Lippincott, Williams & Wilkins. The Journal's web site is located at http://journals.lww.com/spinejournal/pages/default.aspx | - |
dc.relation.ispartof | Spine | - |
dc.rights | This is a non-final version of an article published in final form in Spine, 2020, v. 45 n. 6, p. E346-E351 | - |
dc.subject | Complication | - |
dc.subject | Deformed threads | - |
dc.subject | Distraction failure mechanism | - |
dc.subject | Magnetically controlled growing rod | - |
dc.subject | Rod slippage | - |
dc.title | The Crooked Rod Sign: A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure | - |
dc.type | Article | - |
dc.identifier.email | Cheung, JPY: cheungjp@hku.hk | - |
dc.identifier.email | Zhang, T: tgzhang@hku.hk | - |
dc.identifier.email | Bow, HYC: cbow@hku.hk | - |
dc.identifier.email | Kwan, KYH: kyhkwan@hku.hk | - |
dc.identifier.email | Sze, KY: kysze@hku.hk | - |
dc.identifier.email | Cheung, KMC: cheungmc@hku.hk | - |
dc.identifier.authority | Cheung, JPY=rp01685 | - |
dc.identifier.authority | Kwan, KYH=rp02014 | - |
dc.identifier.authority | Sze, KY=rp00171 | - |
dc.identifier.authority | Cheung, KMC=rp00387 | - |
dc.description.nature | postprint | - |
dc.identifier.doi | 10.1097/BRS.0000000000003268 | - |
dc.identifier.pmid | 31574056 | - |
dc.identifier.scopus | eid_2-s2.0-85080835741 | - |
dc.identifier.hkuros | 308652 | - |
dc.identifier.volume | 45 | - |
dc.identifier.issue | 6 | - |
dc.identifier.spage | E346 | - |
dc.identifier.epage | E351 | - |
dc.identifier.isi | WOS:000544932100007 | - |
dc.publisher.place | United States | - |
dc.identifier.issnl | 0362-2436 | - |