File Download
There are no files associated with this item.
Links for fulltext
(May Require Subscription)
- Publisher Website: 10.1002/rcs.1949
- Scopus: eid_2-s2.0-85052824383
- PMID: 30152081
- WOS: WOS:000449691300007
- Find via
Supplementary
- Citations:
- Appears in Collections:
Article: Intraocular robotic interventional surgical system (IRISS): Semi-automated OCT-guided cataract removal
Title | Intraocular robotic interventional surgical system (IRISS): Semi-automated OCT-guided cataract removal |
---|---|
Authors | |
Keywords | automated cataract surgery cataract extraction intraocular surgery OCT-guided surgery robot-assisted ophthalmic surgery robotic surgery |
Issue Date | 2018 |
Citation | International Journal of Medical Robotics and Computer Assisted Surgery, 2018, v. 14, n. 6, article no. e1949 How to Cite? |
Abstract | Background: With the development of laser-assisted platforms, the outcomes of cataract surgery have been improved by automating several procedures. The cataract-extraction step continues to be manually performed, but due to deficiencies in sensing capabilities, surgical complications such as posterior capsule rupture and incomplete cataract removal remain. Methods: An optical coherence tomography (OCT) system is integrated into our intraocular robotic interventional surgical system (IRISS) robot. The OCT images are used for preoperative planning and intraoperative intervention in a series of automated procedures. Real-time intervention allows surgeons to evaluate the progress and override the operation. Results: The developed system was validated by performing lens extraction on 30 postmortem pig eyes. Complete lens extraction was achieved on 25 eyes, and “almost complete” extraction was achieved on the remainder due to an inability to image small lens particles behind the iris. No capsule rupture was found. Conclusion: The IRISS successfully demonstrated semiautomated OCT-guided lens removal with real-time supervision and intervention. |
Persistent Identifier | http://hdl.handle.net/10722/325413 |
ISSN | 2023 Impact Factor: 2.3 2023 SCImago Journal Rankings: 0.589 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Chen, Cheng Wei | - |
dc.contributor.author | Lee, Yu Hsiu | - |
dc.contributor.author | Gerber, Matthew J. | - |
dc.contributor.author | Cheng, Harrison | - |
dc.contributor.author | Yang, Yan Chao | - |
dc.contributor.author | Govetto, Andrea | - |
dc.contributor.author | Francone, Anibal Andrés | - |
dc.contributor.author | Soatto, Stefano | - |
dc.contributor.author | Grundfest, Warren S. | - |
dc.contributor.author | Hubschman, Jean Pierre | - |
dc.contributor.author | Tsao, Tsu Chin | - |
dc.date.accessioned | 2023-02-27T07:32:39Z | - |
dc.date.available | 2023-02-27T07:32:39Z | - |
dc.date.issued | 2018 | - |
dc.identifier.citation | International Journal of Medical Robotics and Computer Assisted Surgery, 2018, v. 14, n. 6, article no. e1949 | - |
dc.identifier.issn | 1478-5951 | - |
dc.identifier.uri | http://hdl.handle.net/10722/325413 | - |
dc.description.abstract | Background: With the development of laser-assisted platforms, the outcomes of cataract surgery have been improved by automating several procedures. The cataract-extraction step continues to be manually performed, but due to deficiencies in sensing capabilities, surgical complications such as posterior capsule rupture and incomplete cataract removal remain. Methods: An optical coherence tomography (OCT) system is integrated into our intraocular robotic interventional surgical system (IRISS) robot. The OCT images are used for preoperative planning and intraoperative intervention in a series of automated procedures. Real-time intervention allows surgeons to evaluate the progress and override the operation. Results: The developed system was validated by performing lens extraction on 30 postmortem pig eyes. Complete lens extraction was achieved on 25 eyes, and “almost complete” extraction was achieved on the remainder due to an inability to image small lens particles behind the iris. No capsule rupture was found. Conclusion: The IRISS successfully demonstrated semiautomated OCT-guided lens removal with real-time supervision and intervention. | - |
dc.language | eng | - |
dc.relation.ispartof | International Journal of Medical Robotics and Computer Assisted Surgery | - |
dc.subject | automated cataract surgery | - |
dc.subject | cataract extraction | - |
dc.subject | intraocular surgery | - |
dc.subject | OCT-guided surgery | - |
dc.subject | robot-assisted ophthalmic surgery | - |
dc.subject | robotic surgery | - |
dc.title | Intraocular robotic interventional surgical system (IRISS): Semi-automated OCT-guided cataract removal | - |
dc.type | Article | - |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1002/rcs.1949 | - |
dc.identifier.pmid | 30152081 | - |
dc.identifier.scopus | eid_2-s2.0-85052824383 | - |
dc.identifier.volume | 14 | - |
dc.identifier.issue | 6 | - |
dc.identifier.spage | article no. e1949 | - |
dc.identifier.epage | article no. e1949 | - |
dc.identifier.eissn | 1478-596X | - |
dc.identifier.isi | WOS:000449691300007 | - |