Conference Paper: Improved SENSE reconstruction by pre-scaled TLS method

TitleImproved SENSE reconstruction by pre-scaled TLS method
Authors
Issue Date2005
PublisherInternational Society for Magnetic Resonance in Medicine.
Citation
International Society for Magnetic Resonance in Medicine 13th Scientific Meeting & Exhibition, Miami, FL, 7-13 May 2005, p. 2440 How to Cite?
AbstractIn SENSE reconstruction, the unfolding process usually involves solving an over-determined linear system of equations (LSE) using a least squares (LS) estimation. In this work it is proposed that when the coil sensitivity maps can not be accurately calibrated the pre-scaled total least squares (TLS) method is more suitable for unfolding aliased pixels. By carefully choosing the relative scaling parameter for the LSE, the TLS method leads to more accurate estimation than ordinary LS method and hence the reconstructed images has less aliasing artifacts and better SNR.
Persistent Identifierhttp://hdl.handle.net/10722/99028

 

DC FieldValueLanguage
dc.contributor.authorQu, Pen_HK
dc.contributor.authorYuan, Jen_HK
dc.contributor.authorWei, Jen_HK
dc.contributor.authorShen, GGen_HK
dc.date.accessioned2010-09-25T18:12:56Z-
dc.date.available2010-09-25T18:12:56Z-
dc.date.issued2005en_HK
dc.identifier.citationInternational Society for Magnetic Resonance in Medicine 13th Scientific Meeting & Exhibition, Miami, FL, 7-13 May 2005, p. 2440en_HK
dc.identifier.urihttp://hdl.handle.net/10722/99028-
dc.description.abstractIn SENSE reconstruction, the unfolding process usually involves solving an over-determined linear system of equations (LSE) using a least squares (LS) estimation. In this work it is proposed that when the coil sensitivity maps can not be accurately calibrated the pre-scaled total least squares (TLS) method is more suitable for unfolding aliased pixels. By carefully choosing the relative scaling parameter for the LSE, the TLS method leads to more accurate estimation than ordinary LS method and hence the reconstructed images has less aliasing artifacts and better SNR.-
dc.languageengen_HK
dc.publisherInternational Society for Magnetic Resonance in Medicine.en_HK
dc.relation.ispartofInternational Society for Magnetic Resonance in Medicine Scientific Meeting & Exhibitionen_HK
dc.titleImproved SENSE reconstruction by pre-scaled TLS methoden_HK
dc.typeConference_Paperen_HK
dc.identifier.emailShen, GG: gxshen@eee.hku.hken_HK
dc.identifier.authorityShen, GG=rp00166en_HK
dc.identifier.hkuros104538en_HK
dc.identifier.volume, v. 1en_HK
dc.identifier.spage2440en_HK