Conference Paper: Comparison of B1 Field Homogeneity for Shielded Birdcage, TEM and Microstrip Volume Coils at 300MHz

TitleComparison of B1 Field Homogeneity for Shielded Birdcage, TEM and Microstrip Volume Coils at 300MHz
Authors
Issue Date2004
PublisherThe International Society for Magnetic Resonance in Medicine (ISMRM)
Citation
The International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 487 How to Cite?
AbstractIn this work the B1 fields of shielded birdcage, TEM and microstrip volume coils were simulated by the finite difference time domain (FDTD) method at 300MHz under the same geometry conditions. The simulation results indicate that the shielded birdcage coil has better B1 field homogeneity compared with the TEM and microstrip volume coils, but might have more radiation loss than the TEM and Microstrip volume coils.
Persistent Identifierhttp://hdl.handle.net/10722/99215
ISSN

 

DC FieldValueLanguage
dc.contributor.authorWang, Cen_HK
dc.contributor.authorYuan, Jen_HK
dc.contributor.authorWu, Ben_HK
dc.contributor.authorShen, GGen_HK
dc.date.accessioned2010-09-25T18:20:32Z-
dc.date.available2010-09-25T18:20:32Z-
dc.date.issued2004en_HK
dc.identifier.citationThe International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 487en_HK
dc.identifier.issn1545-4436-
dc.identifier.urihttp://hdl.handle.net/10722/99215-
dc.description.abstractIn this work the B1 fields of shielded birdcage, TEM and microstrip volume coils were simulated by the finite difference time domain (FDTD) method at 300MHz under the same geometry conditions. The simulation results indicate that the shielded birdcage coil has better B1 field homogeneity compared with the TEM and microstrip volume coils, but might have more radiation loss than the TEM and Microstrip volume coils.-
dc.languageengen_HK
dc.publisherThe International Society for Magnetic Resonance in Medicine (ISMRM)en_HK
dc.relation.ispartofProceedings of the International Society for Magnetic Resonance in Medicine (ISMRM)en_HK
dc.titleComparison of B1 Field Homogeneity for Shielded Birdcage, TEM and Microstrip Volume Coils at 300MHzen_HK
dc.typeConference_Paperen_HK
dc.identifier.emailShen, GG: gxshen@eee.hku.hken_HK
dc.identifier.authorityShen, GG=rp00166en_HK
dc.description.naturepublished_or_final_version-
dc.identifier.hkuros91309en_HK
dc.identifier.volume11en_HK
dc.identifier.spage487en_HK
dc.identifier.issnl1524-6965-