File Download
Supplementary
-
Citations:
- Appears in Collections:
Conference Paper: Comparison of B1 Field Homogeneity for Shielded Birdcage, TEM and Microstrip Volume Coils at 300MHz
Title | Comparison of B1 Field Homogeneity for Shielded Birdcage, TEM and Microstrip Volume Coils at 300MHz |
---|---|
Authors | |
Issue Date | 2004 |
Publisher | The International Society for Magnetic Resonance in Medicine (ISMRM) |
Citation | The International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 487 How to Cite? |
Abstract | In this work the B1 fields of shielded birdcage, TEM and microstrip volume coils were simulated by the finite difference time domain (FDTD) method at
300MHz under the same geometry conditions. The simulation results indicate that the shielded birdcage coil has better B1 field homogeneity compared with
the TEM and microstrip volume coils, but might have more radiation loss than the TEM and Microstrip volume coils. |
Persistent Identifier | http://hdl.handle.net/10722/99215 |
ISSN |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Wang, C | en_HK |
dc.contributor.author | Yuan, J | en_HK |
dc.contributor.author | Wu, B | en_HK |
dc.contributor.author | Shen, GG | en_HK |
dc.date.accessioned | 2010-09-25T18:20:32Z | - |
dc.date.available | 2010-09-25T18:20:32Z | - |
dc.date.issued | 2004 | en_HK |
dc.identifier.citation | The International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 487 | en_HK |
dc.identifier.issn | 1545-4436 | - |
dc.identifier.uri | http://hdl.handle.net/10722/99215 | - |
dc.description.abstract | In this work the B1 fields of shielded birdcage, TEM and microstrip volume coils were simulated by the finite difference time domain (FDTD) method at 300MHz under the same geometry conditions. The simulation results indicate that the shielded birdcage coil has better B1 field homogeneity compared with the TEM and microstrip volume coils, but might have more radiation loss than the TEM and Microstrip volume coils. | - |
dc.language | eng | en_HK |
dc.publisher | The International Society for Magnetic Resonance in Medicine (ISMRM) | en_HK |
dc.relation.ispartof | Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM) | en_HK |
dc.title | Comparison of B1 Field Homogeneity for Shielded Birdcage, TEM and Microstrip Volume Coils at 300MHz | en_HK |
dc.type | Conference_Paper | en_HK |
dc.identifier.email | Shen, GG: gxshen@eee.hku.hk | en_HK |
dc.identifier.authority | Shen, GG=rp00166 | en_HK |
dc.description.nature | published_or_final_version | - |
dc.identifier.hkuros | 91309 | en_HK |
dc.identifier.volume | 11 | en_HK |
dc.identifier.spage | 487 | en_HK |
dc.identifier.issnl | 1524-6965 | - |